CVE-2014-5434

No exploitation reported by the sources we read.

  • Published2019-03-26
  • WeaknessCWE-259 · CWE-798

Am I affected?

With a domain, we read the version its public headers and pages expose. A distribution build can carry the fix under an older number.

Description

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 has a default account with hard-coded credentials used with the FTP protocol. Baxter asserts no files can be transferred to or from the WBM using this account. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.

Timeline

dates in UTC
  1. 2019-03-2615:29CVE publishedNVD

Observed activity

isMalicious corpus · updated 2026-10-23 18:48 UTC
Indicators tagged with this CVE, per day0 indicators over 7 days
Block these indicators
  • Firewall list (EDL, text): https://api.ismalicious.com/edl/cve/CVE-2014-5434?format=txt
  • CSV export: https://api.ismalicious.com/edl/cve/CVE-2014-5434?format=csv
  • TAXII collection of exploited CVEs
  • The EDL and CSV take an API key on the Pro plan.

Versions per branch

NVD
Baxter sigma spectrum infusion system firmwareversion to verify
BranchAffectedFixed in
6.5= 6.05no fix published

as publishedRanges as NVD and OSV publish them; a distribution’s build can be fixed under an older number.

Distribution builds and vendor advisories

Packages from OSV.dev and products from the vendors' CSAF advisories.

version not verifiedVersions not verified: read as OSV.dev and the vendors publish them, not checked against each vendor's release list.

Vendor advisories

  • ICSA-15-181-01 CISA1 known affected

    Baxter SIGMA Spectrum Infusion System Vulnerabilities

    Baxter SIGMA Spectrum Infusion System(model 35700BAX): 6.05

Full record on NVD

Fix

  1. Mitigation from the vendor CISA ICSA-15-181-01 · 1 productBaxter offers the following recommendations to help mitigate risks associated with these vulnerabilities in the SIGMA Spectrum Infusion System running Version 6.05 with WBM Version 16
  2. Mitigation from the vendor CISA ICSA-15-181-01 · 1 productEnsure that the WI-FI network supporting WBMs is secured using a secure WI-FI protocol
  3. Mitigation from the vendor CISA ICSA-15-181-01 · 1 productSeparate the network supporting the WBMs with a standalone VLAN or use similarly segmented network topography to isolate WBMs. This would require an attacker to compromise the standalone WI-FI network or otherwise gain access to the supporting VLAN before SSH access to the WBM is possible
  4. Mitigation from the vendor CISA ICSA-15-181-01 · 1 productConfigure Wireless Access Points and Firewalls, which provide access to the VLAN, to block Port 21/FTP and Port 22/SSH
  5. Mitigation from the vendor CISA ICSA-15-181-01 · 1 productEnsure that network authentication credentials used by the WBM to connect to the network are properly restricted to only allow access to the wireless network
  6. Mitigation from the vendor CISA ICSA-15-181-01 · 1 productAs a last resort, customers may disable wireless operation of the pump. The Sigma Spectrum Infusion System was designed to operate without network access. This action would impact an organization’s ability to rapidly deploy drug library (formulary) updates to their pumps

Advisories and patches

Detection

Nuclei template
none so far
ProjectDiscovery
Exploit-DB
no entry
Exploit-DB
Proof-of-concept repositories
none listed
PoC-in-GitHub

Sources

  • NVDCVSS 3.09.8read 2026-10-23
  • NVDCVSS 2.05read 2026-10-23
  • CNAweakness (CWE)CWE-259read 2026-10-23
  • NVDweakness (CWE)CWE-798read 2026-10-23
  • FIRST EPSSEPSS score1.6 %2026-10-22

Technical details

CVSS v3 Vector
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Published
2019-03-26
Last Modified
2026-06-17
NVD status
Modified

Frequently asked questions

What is CVE-2014-5434?

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 has a default account with hard-coded credentials used with the FTP protocol. Baxter asserts no files can be transferred to or from the WBM using this account. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.

Is CVE-2014-5434 actively exploited?

Active exploitation of CVE-2014-5434 has not been confirmed. Its EPSS score was 1.6% on 2026-10-22, the estimated probability of exploitation in the next 30 days.

What is the CVSS score for CVE-2014-5434?

CVE-2014-5434 has a CVSS v3 base score of 9.8 (CRITICAL severity), with vector string CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H.

Hear about the next vulnerability in your software

Surveillance reads the versions in your stack — a domain, a lockfile or a list — checks each CVE against the ranges its sources publish, and alerts you when one you run is exploited or enters CISA KEV. A version it cannot read stays « version not verified ».

No credit card required · No monthly cap on website lookups · 250 API requests/month