Skip to main content
No exploitation reported

CVE-2025-0683

No exploitation reported by the sources we read.

  • Published2025-01-30
  • WeaknessCWE-359

Am I affected?

With a domain, we read the version its public headers and pages expose. A distribution build can carry the fix under an older number.

Description

In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.

Timeline

dates in UTC
  1. 2025-01-3019:15CVE publishedNVD

Versions per branch

Products declared by the CNA

As icscert, the CNA that assigned this CVE, declared them in the CVE record; versions as written. NVD lists no product for this CVE yet: this is the only product data so far.

  • Contec Health CMS8000 Patient Monitorother versions: not affected
    Affected
    • = All versionsno fix stated

Source: the CNA record (CVE Program), © The MITRE Corporation,CVE Terms of UseCVE recordupdated 2025-01-31

Distribution builds and vendor advisories

Packages from OSV.dev and products from the vendors' CSAF advisories.

version not verifiedVersions not verified: read as OSV.dev and the vendors publish them, not checked against each vendor's release list.

Vendor advisories

  • ICSMA-25-030-01 CISA4 known affected

    Contec Health CMS8000 Patient Monitor (Update A)

    Contec Health CMS8000 Patient Monitor Firmware: smart3250-2.6.27-wlan2.1.7.cramfs · Contec Health CMS8000 Patient Monitor Firmware: CMS7.820.075.08/0.74(0.75) · Contec Health CMS8000 Patient Monitor Firmware: CMS7.820.120.01/0.93(0.95) · +1 more

Full record on NVD

Fix

  1. Mitigation from the vendor CISA ICSMA-25-030-01 · 4 productsPer FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.
  2. Mitigation from the vendor CISA ICSMA-25-030-01 · 4 productsIf asset owners cannot remove the devices from their networks, users should block 202.114.4.0/24 from their networks, or block 202.114.4.119 and 202.114.4.120.
  3. Mitigation from the vendor CISA ICSMA-25-030-01 · 4 productsPlease note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA's safety communication.

Detection

Nuclei template
none so far
ProjectDiscovery
Exploit-DB
no entry
Exploit-DB
Proof of concept
Metasploit modules
none listed
Metasploit Framework

Sources

  • NVDCVE recordread 2026-10-23
  • CISA-ADPSSVC decisionnone2025-01-30
  • CNACVSS 4.08.2known since 2026-10-23
  • CNACVSS 3.15.9known since 2026-10-23
  • CNAweakness (CWE)CWE-359known since 2026-10-23
  • FIRST EPSSEPSS score0.8 %2026-11-18
  • GitHub Advisoriesadvisory—

Identifiers

Technical details

CVSS v3 Vector
CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N
CVSS 4.0 vector
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
Published
2025-01-30
Last Modified
2026-06-17
NVD status
Deferred
GitHub Advisory
GHSA-h39w-8pq5-8qh2

Frequently asked questions

What is CVE-2025-0683?

In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.

Is CVE-2025-0683 actively exploited?

Active exploitation of CVE-2025-0683 has not been confirmed. Its EPSS score was 0.8% on 2026-11-18, the estimated probability of exploitation in the next 30 days.

What is the CVSS score for CVE-2025-0683?

CVE-2025-0683 has a CVSS v3 base score of 5.9 (MEDIUM severity), with vector string CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N.

Running Contec Health CMS8000 Patient? Hear about its next vulnerability

Surveillance reads the versions in your stack — a domain, a lockfile or a list — checks each CVE against the ranges its sources publish, and alerts you when one you run is exploited or enters CISA KEV. A version it cannot read stays « version not verified ».

No credit card required · No monthly cap on website lookups · 250 API requests/month

Sources for this record