CVE-2025-0683
No exploitation reported by the sources we read.
- Published2025-01-30
- WeaknessCWE-359
Am I affected?
Debian, Ubuntu or Red Hat sometimes backport a fix without changing the upstream version number: such a version reads « to verify », never « affected ».With a domain, we read the version its public headers and pages expose. A distribution build can carry the fix under an older number.
Description
In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.
Timeline
dates in UTC- 2025-01-3019:15CVE publishedNVD
Versions per branch
Products declared by the CNA
As icscert, the CNA that assigned this CVE, declared them in the CVE record; versions as written. NVD lists no product for this CVE yet: this is the only product data so far.
- Contec Health CMS8000 Patient Monitorother versions: not affectedAffected
- = All versionsno fix stated
Source: the CNA record (CVE Program), © The MITRE Corporation,CVE Terms of UseCVE recordupdated 2025-01-31
Distribution builds and vendor advisories
Packages from OSV.dev and products from the vendors' CSAF advisories.
version not verifiedVersions not verified: read as OSV.dev and the vendors publish them, not checked against each vendor's release list.
Vendor advisories
Contec Health CMS8000 Patient Monitor (Update A)
Contec Health CMS8000 Patient Monitor Firmware: smart3250-2.6.27-wlan2.1.7.cramfs · Contec Health CMS8000 Patient Monitor Firmware: CMS7.820.075.08/0.74(0.75) · Contec Health CMS8000 Patient Monitor Firmware: CMS7.820.120.01/0.93(0.95) · +1 more
Fix
- Mitigation from the vendor CISA ICSMA-25-030-01 · 4 productsPer FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.
- Mitigation from the vendor CISA ICSMA-25-030-01 · 4 productsIf asset owners cannot remove the devices from their networks, users should block 202.114.4.0/24 from their networks, or block 202.114.4.119 and 202.114.4.120.
- Mitigation from the vendor CISA ICSMA-25-030-01 · 4 productsPlease note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA's safety communication.
Detection
- Nuclei template
- none so far ProjectDiscovery
- Exploit-DB
- no entry Exploit-DB
- Proof of concept
- PoC-in-GitHubno public repository identifiedread 2026-11-19
- Metasploit modules
- none listed Metasploit Framework
Sources
- NVDCVE recordread 2026-10-23
- CISA-ADPSSVC decisionnone2025-01-30
- CNACVSS 4.08.2known since 2026-10-23
- CNACVSS 3.15.9known since 2026-10-23
- CNAweakness (CWE)CWE-359known since 2026-10-23
- FIRST EPSSEPSS score0.8 %2026-11-18
- GitHub Advisoriesadvisory—
Identifiers
- EUVD-2025-1815same flawENISA EUVD · LicenseRef-ENISA-legal-notice
- GHSA-h39w-8pq5-8qh2same flawGitHub Advisory Database · CC-BY-4.0
Technical details
- CVSS v3 Vector
- CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N
- CVSS 4.0 vector
- CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
- Published
- 2025-01-30
- Last Modified
- 2026-06-17
- NVD status
- Deferred
- GitHub Advisory
- GHSA-h39w-8pq5-8qh2
Frequently asked questions
What is CVE-2025-0683?
Is CVE-2025-0683 actively exploited?
What is the CVSS score for CVE-2025-0683?
Running Contec Health CMS8000 Patient? Hear about its next vulnerability
Surveillance reads the versions in your stack — a domain, a lockfile or a list — checks each CVE against the ranges its sources publish, and alerts you when one you run is exploited or enters CISA KEV. A version it cannot read stays « version not verified ».
No credit card required · No monthly cap on website lookups · 250 API requests/month
Go further
- Following what attackers exploit right now? The CISA KEV additions of the month, with their patch deadlines.
- Running this software? Vulnerability monitoring alerts you to new vulnerabilities in your products.
Related vulnerabilities
Same product first, then the same vendor advisory and the same publication week.
Sources for this record
- CISA ICS advisories (CSAF) · Public domain (US Government work)
- CVE Program · CVE Program Terms of Use
- ENISA EU Vulnerability Database · ENISA legal notice §1.2 (reproduction authorised, provided the source is acknowledged)
- FIRST EPSS · EPSS terms of use (free use, attribution requested)
- GitHub Advisory Database · CC-BY-4.0 · GHSA-h39w-8pq5-8qh2
- NVD · Public domain (17 U.S.C. § 105) for NVD-authored data; third-party content under its own terms
- CISA Vulnrichment · CC0-1.0