Skip to main content
No exploitation reported

CVE-2024-12248

No exploitation reported by the sources we read.

  • Published2025-01-30
  • WeaknessCWE-787

Am I affected?

With a domain, we read the version its public headers and pages expose. A distribution build can carry the fix under an older number.

Description

Contec Health CMS8000 Patient Monitor is vulnerable to an out-of-bounds write, which could allow an attacker to send specially formatted UDP requests in order to write arbitrary data. This could result in remote code execution.

Timeline

dates in UTC
  1. 2025-01-3019:15CVE publishedNVD

Versions per branch

Products declared by the CNA

As icscert, the CNA that assigned this CVE, declared them in the CVE record; versions as written. NVD lists no product for this CVE yet: this is the only product data so far.

  • Contec Health CMS8000 Patient Monitorother versions: not affected
    Affected
    • = Firmware version smart3250-2.6.27-wlan2.1.7.cramfsno fix stated
    • = Firmware version CMS7.820.075.08/0.74(0.75)no fix stated
    • = Firmware version CMS7.820.120.01/0.93(0.95)no fix stated

Source: the CNA record (CVE Program), © The MITRE Corporation,CVE Terms of UseCVE recordupdated 2025-01-31

Distribution builds and vendor advisories

Packages from OSV.dev and products from the vendors' CSAF advisories.

version not verifiedVersions not verified: read as OSV.dev and the vendors publish them, not checked against each vendor's release list.

Vendor advisories

  • ICSMA-25-030-01 CISA3 known affected

    Contec Health CMS8000 Patient Monitor (Update A)

    Contec Health CMS8000 Patient Monitor Firmware: smart3250-2.6.27-wlan2.1.7.cramfs · Contec Health CMS8000 Patient Monitor Firmware: CMS7.820.075.08/0.74(0.75) · Contec Health CMS8000 Patient Monitor Firmware: CMS7.820.120.01/0.93(0.95)

Full record on NVD

Fix

  1. Mitigation from the vendor CISA ICSMA-25-030-01 · 3 productsPer FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.
  2. Mitigation from the vendor CISA ICSMA-25-030-01 · 3 productsIf asset owners cannot remove the devices from their networks, users should block 202.114.4.0/24 from their networks, or block 202.114.4.119 and 202.114.4.120.
  3. Mitigation from the vendor CISA ICSMA-25-030-01 · 3 productsPlease note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA's safety communication.

Detection

Nuclei template
none so far
ProjectDiscovery
Exploit-DB
no entry
Exploit-DB
Proof of concept
Metasploit modules
none listed
Metasploit Framework

Sources

  • NVDCVE recordread 2026-10-23
  • CISA-ADPSSVC decisionnone2025-01-31
  • CNACVSS 4.09.3known since 2026-10-23
  • CNACVSS 3.19.8known since 2026-10-23
  • CNAweakness (CWE)CWE-787known since 2026-10-23
  • FIRST EPSSEPSS score1.3 %2026-11-18
  • GitHub Advisoriesadvisory—

Identifiers

Technical details

CVSS v3 Vector
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
CVSS 4.0 vector
CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
Published
2025-01-30
Last Modified
2026-06-17
NVD status
Deferred
GitHub Advisory
GHSA-w5vx-f3xq-8wcg

Frequently asked questions

What is CVE-2024-12248?

Contec Health CMS8000 Patient Monitor is vulnerable to an out-of-bounds write, which could allow an attacker to send specially formatted UDP requests in order to write arbitrary data. This could result in remote code execution.

Is CVE-2024-12248 actively exploited?

Active exploitation of CVE-2024-12248 has not been confirmed. Its EPSS score was 1.3% on 2026-11-18, the estimated probability of exploitation in the next 30 days.

What is the CVSS score for CVE-2024-12248?

CVE-2024-12248 has a CVSS v3 base score of 9.8 (CRITICAL severity), with vector string CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H.

Running Contec Health CMS8000 Patient? Hear about its next vulnerability

Surveillance reads the versions in your stack — a domain, a lockfile or a list — checks each CVE against the ranges its sources publish, and alerts you when one you run is exploited or enters CISA KEV. A version it cannot read stays « version not verified ».

No credit card required · No monthly cap on website lookups · 250 API requests/month

Sources for this record