Skip to main content
No exploitation reported

CVE-2025-0626

No exploitation reported by the sources we read.

  • Published2025-01-30
  • WeaknessCWE-912

Am I affected?

With a domain, we read the version its public headers and pages expose. A distribution build can carry the fix under an older number.

Description

The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.

Timeline

dates in UTC
  1. 2025-01-3019:15CVE publishedNVD

Versions per branch

Products declared by the CNA

As icscert, the CNA that assigned this CVE, declared them in the CVE record; versions as written. NVD lists no product for this CVE yet: this is the only product data so far.

  • Contec Health CMS8000 Patient Monitorother versions: not affected
    Affected
    • = All versionsno fix stated

Source: the CNA record (CVE Program), © The MITRE Corporation,CVE Terms of UseCVE recordupdated 2025-03-01

Distribution builds and vendor advisories

Packages from OSV.dev and products from the vendors' CSAF advisories.

version not verifiedVersions not verified: read as OSV.dev and the vendors publish them, not checked against each vendor's release list.

Vendor advisories

  • ICSMA-25-030-01 CISA4 known affected

    Contec Health CMS8000 Patient Monitor (Update A)

    Contec Health CMS8000 Patient Monitor Firmware: smart3250-2.6.27-wlan2.1.7.cramfs · Contec Health CMS8000 Patient Monitor Firmware: CMS7.820.075.08/0.74(0.75) · Contec Health CMS8000 Patient Monitor Firmware: CMS7.820.120.01/0.93(0.95) · +1 more

Full record on NVD

Fix

  1. Mitigation from the vendor CISA ICSMA-25-030-01 · 4 productsPer FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.
  2. Mitigation from the vendor CISA ICSMA-25-030-01 · 4 productsIf asset owners cannot remove the devices from their networks, users should block 202.114.4.0/24 from their networks, or block 202.114.4.119 and 202.114.4.120.
  3. Mitigation from the vendor CISA ICSMA-25-030-01 · 4 productsPlease note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA's safety communication.

Detection

Nuclei template
none so far
ProjectDiscovery
Exploit-DB
no entry
Exploit-DB
Proof of concept
Metasploit modules
none listed
Metasploit Framework

Sources

  • NVDCVE recordread 2026-10-23
  • CISA-ADPSSVC decisionnone2025-01-30
  • CNACVSS 4.07.7known since 2026-10-23
  • CNACVSS 3.17.5known since 2026-10-23
  • CNAweakness (CWE)CWE-912known since 2026-10-23
  • FIRST EPSSEPSS score1.1 %2026-11-18
  • GitHub Advisoriesadvisory—

Identifiers

Technical details

CVSS v3 Vector
CVSS:3.1/AV:N/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H
CVSS 4.0 vector
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
Published
2025-01-30
Last Modified
2026-06-17
NVD status
Deferred
GitHub Advisory
GHSA-p2j3-fm26-85wv

Frequently asked questions

What is CVE-2025-0626?

The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.

Is CVE-2025-0626 actively exploited?

Active exploitation of CVE-2025-0626 has not been confirmed. Its EPSS score was 1.1% on 2026-11-18, the estimated probability of exploitation in the next 30 days.

What is the CVSS score for CVE-2025-0626?

CVE-2025-0626 has a CVSS v3 base score of 7.5 (HIGH severity), with vector string CVSS:3.1/AV:N/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H.