Debian, Ubuntu or Red Hat sometimes backport a fix without changing the upstream version number: such a version reads « to verify », never « affected ».With a domain, we read the version its public headers and pages expose. A distribution build can carry the fix under an older number.
Description
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensitive data stored on the WBM that permits temporary configuration changes to network settings of the WBM, and allows the WBM to be rebooted. Temporary configuration changes to network settings are removed upon reboot.
Timeline
dates in UTC- 2020-06-2914:15CVE publishedNVD
Observed activity
isMalicious corpus · updated 2026-10-23 21:37 UTCBlock these indicators
- Firewall list (EDL, text): https://api.ismalicious.com/edl/cve/CVE-2020-12041?format=txt
- CSV export: https://api.ismalicious.com/edl/cve/CVE-2020-12041?format=csv
- TAXII collection of exploited CVEs
- The EDL and CSV take an API key on the Pro plan.
Versions per branch
NVD| Branch | Affected | Fixed in | Source |
|---|---|---|---|
| 8.0 | = 8.0 | no fix published | NVD |
as publishedRanges as NVD and OSV publish them; a distribution’s build can be fixed under an older number.
Distribution builds and vendor advisories
Packages from OSV.dev and products from the vendors' CSAF advisories.
version not verifiedVersions not verified: read as OSV.dev and the vendors publish them, not checked against each vendor's release list.
Vendor advisories
Baxter Sigma Spectrum Infusion Pumps (Update B)
Baxter Spectrum: v8.x model 35700BAX2 · Sigma Spectrum LVP: v6.x with Wireless Battery Modules v9 v11 v13 v14 v15 v16 v16D38 v17 v17D19 v20D29 to v20D32 and v22D24 to v22D28 · Sigma Spectrum: v6.x model 35700BAX · +3 more
Fix
- Vendor fix CISA ICSMA-20-170-04 · 6 productsBaxter recommends ensuring appropriate physical controls within user environments to protect against unauthorized access to devices.
- Vendor fix CISA ICSMA-20-170-04 · 6 productsBaxter recommends isolating the Spectrum Infusion Systems to its own network VLAN to segregate the system from other hospital systems, and reduce the probability that a threat actor could execute an adjacent attack such as a MiTM attack against the system to observe clear-text communications.
- Vendor fix CISA ICSMA-20-170-04 · 6 productsBaxter recommends using the strongest available wireless network security protocols (e.g., WPA2, EAP-TLS, etc.) to provide authentication and encryption of wireless data sent to and from the Spectrum Infusion System.
- Vendor fix CISA ICSMA-20-170-04 · 6 productsUsers should ensure the WBM is rebooted after configuration for their network(s) by removing the WBM from the rear of the Spectrum device for 10-15 seconds, and then re-attaching the WBM.
- Vendor fix CISA ICSMA-20-170-04 · 6 productsUsers should always monitor for and/or block unexpected traffic, such as FTP, at network boundaries into the Spectrum-specific VLAN.
- Vendor fix CISA ICSMA-20-170-04 · 6 productsAs a last resort, users may disable wireless operation of the pump. The Spectrum Infusion System was designed to operate without network access. This action would impact an organization 's ability to rapidly deploy drug library (formulary) updates to their pumps.
Detection
- Nuclei template
- none so far ProjectDiscovery
- Exploit-DB
- no entry Exploit-DB
- Proof-of-concept repositories
- none listed PoC-in-GitHub
Sources
- NVDCVSS 3.19.4read 2026-10-23
- NVDCVSS 2.07.5read 2026-10-23
- CNAweakness (CWE)CWE-732read 2026-10-23
- NVDweakness (CWE)CWE-732read 2026-10-23
- FIRST EPSSEPSS score1.4 %2026-10-22
Technical details
- CVSS v3 Vector
- CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:L/A:H
- Published
- 2020-06-29
- Last Modified
- 2026-06-17
- NVD status
- Modified
Frequently asked questions
What is CVE-2020-12041?
Is CVE-2020-12041 actively exploited?
What is the CVSS score for CVE-2020-12041?
Hear about the next vulnerability in your software
Surveillance reads the versions in your stack — a domain, a lockfile or a list — checks each CVE against the ranges its sources publish, and alerts you when one you run is exploited or enters CISA KEV. A version it cannot read stays « version not verified ».
No credit card required · No monthly cap on website lookups · 250 API requests/month
Go further
- Following what attackers exploit right now? The CISA KEV additions of the month, with their patch deadlines.
- Running this software? Vulnerability monitoring alerts you to new vulnerabilities in your products.
Related vulnerabilities
Same product first, then the same vendor advisory and the same publication week.