Skip to main content
No exploitation reportedFix available

CVE-2019-13530

No exploitation reported by the sources we read.

  • Published2019-09-12
  • WeaknessCWE-259 · CWE-798

Am I affected?

With a domain, we read the version its public headers and pages expose. A distribution build can carry the fix under an older number.

Description

Philips IntelliVue WLAN, portable patient monitors, WLAN Version A, Firmware A.03.09, WLAN Version A, Firmware A.03.09, Part #: M8096-67501, WLAN Version B, Firmware A.01.09, Part #: N/A (Replaced by Version C) and WLAN Version B, Firmware A.01.09, Part #: N/A (Replaced by Version C). An attacker can use these credentials to login via ftp and upload a malicious firmware.

Timeline

dates in UTC
  1. 2019-09-1220:15CVE publishedNVD

Observed activity

isMalicious corpus · updated 2026-10-23 18:48 UTC
Indicators tagged with this CVE, per day0 indicators over 7 days
Block these indicators
  • Firewall list (EDL, text): https://api.ismalicious.com/edl/cve/CVE-2019-13530?format=txt
  • CSV export: https://api.ismalicious.com/edl/cve/CVE-2019-13530?format=csv
  • TAXII collection of exploited CVEs
  • The EDL and CSV take an API key on the Pro plan.

Versions per branch

NVD
Philips intellivue mp monitors mp2\/x2 firmwareversion to verify
BranchAffectedFixed in
—= a01.09no fix published
Philips intellivue mp monitors mp20-mp90 firmwareversion to verify
BranchAffectedFixed in
—= a.03.09no fix published
Philips intellivue mp monitors mp5\/5sc firmwareversion to verify
BranchAffectedFixed in
—= a.03.09no fix published

1 more product in the sources.

as publishedRanges as NVD and OSV publish them; a distribution’s build can be fixed under an older number.

Distribution builds and vendor advisories

Packages from OSV.dev and products from the vendors' CSAF advisories.

version not verifiedVersions not verified: read as OSV.dev and the vendors publish them, not checked against each vendor's release list.

Vendor advisories

  • ICSMA-19-255-01 CISA4 known affected

    Philips IntelliVue WLAN

    IntelliVue MP monitors MX800/700/600 ((865240/41/42) WLAN: Version B, Firmware A.01.09, Part #: N/A (Replaced by Version C) · IntelliVue MP monitors MP2/X2 (M8102A/M3002A) WLAN: Version B, Firmware A.01.09, Part #: N/A (Replaced by Version C) · IntelliVue MP monitors MP5/5SC (M8105A/5AS) WLAN: Version A, Firmware A.03.09, Part #: M8096-67501 · +1 more

Full record on NVD

Fix

  1. Vendor fix CISA ICSMA-19-255-01 · 4 productsWLAN Version B is obsolete. Philips recommends users update to the WLAN assembly Version C wireless module in affected IntelliVue monitors.
  2. Vendor fix CISA ICSMA-19-255-01 · 4 productsWLAN Version C with current firmware of B.00.31 is not vulnerable to the described attack.
  3. Vendor fix CISA ICSMA-19-255-01 · 4 productsWLAN Version A is a software patch from Philips estimated to be available in Incenter by the end of 2019.
  4. Vendor fix CISA ICSMA-19-255-01 · 4 productsWireless network access should be controlled by authentication and authorization (e.g., WPA2), which are supported by Philips.
  5. Vendor fix CISA ICSMA-19-255-01 · 4 productsAs an additional mitigation measure, Philips recommends implementing a firewall rule on the user 's wireless network to restrict access control via FTP. Users should implement role-based access controls to control physical access to the system. Refer to the Security for Clinical Networks Guide and …
  6. Vendor fix CISA ICSMA-19-255-01 · 4 productsPlease see the Philips product security website for the latest security information for Philips products:https://www.philips.com/productsecurity

Advisories and patches

Detection

Nuclei template
none so far
ProjectDiscovery
Exploit-DB
no entry
Exploit-DB
Proof-of-concept repositories
none listed
PoC-in-GitHub

Sources

  • NVDCVSS 3.17.2read 2026-10-23
  • NVDCVSS 2.06.5read 2026-10-23
  • CNAweakness (CWE)CWE-259read 2026-10-23
  • NVDweakness (CWE)CWE-798read 2026-10-23
  • FIRST EPSSEPSS score1.4 %2026-10-22

Technical details

CVSS v3 Vector
CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H
Published
2019-09-12
Last Modified
2026-06-17
NVD status
Modified

Frequently asked questions

What is CVE-2019-13530?

Philips IntelliVue WLAN, portable patient monitors, WLAN Version A, Firmware A.03.09, WLAN Version A, Firmware A.03.09, Part #: M8096-67501, WLAN Version B, Firmware A.01.09, Part #: N/A (Replaced by Version C) and WLAN Version B, Firmware A.01.09, Part #: N/A (Replaced by Version C). An attacker can use these credentials to login via ftp and upload a malicious firmware.

Is CVE-2019-13530 actively exploited?

Active exploitation of CVE-2019-13530 has not been confirmed. Its EPSS score was 1.4% on 2026-10-22, the estimated probability of exploitation in the next 30 days.

What is the CVSS score for CVE-2019-13530?

CVE-2019-13530 has a CVSS v3 base score of 7.2 (HIGH severity), with vector string CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H.

Running Philips IntelliVue WLAN? Hear about its next vulnerability

Surveillance reads the versions in your stack — a domain, a lockfile or a list — checks each CVE against the ranges its sources publish, and alerts you when one you run is exploited or enters CISA KEV. A version it cannot read stays « version not verified ».

No credit card required · No monthly cap on website lookups · 250 API requests/month