Skip to main content
No exploitation reportedFix available

CVE-2018-7518

No exploitation reported by the sources we read.

  • Published2018-05-24
  • Fixed in4107600010.23
  • WeaknessCWE-522

Am I affected?

With a domain, we read the version its public headers and pages expose. A distribution build can carry the fix under an older number.

Description

In TotalAlert Web Application in BeaconMedaes Scroll Medical Air Systems prior to v4107600010.23, an attacker with network access to the integrated web server could retrieve default or user defined credentials stored and transmitted in an insecure manner.

Timeline

dates in UTC
  1. 2018-05-2420:29CVE publishedNVD

Observed activity

isMalicious corpus · updated 2026-10-23 18:48 UTC
Indicators tagged with this CVE, per day0 indicators over 7 days
Block these indicators
  • Firewall list (EDL, text): https://api.ismalicious.com/edl/cve/CVE-2018-7518?format=txt
  • CSV export: https://api.ismalicious.com/edl/cve/CVE-2018-7518?format=csv
  • TAXII collection of exploited CVEs
  • The EDL and CSV take an API key on the Pro plan.

Versions per branch

NVD
Beaconmedaes scroll medical air systems firmwareversion to verify
BranchAffectedFixed in
4107600010.23< 4107600010.234107600010.23

as publishedRanges as NVD and OSV publish them; a distribution’s build can be fixed under an older number.

Distribution builds and vendor advisories

Packages from OSV.dev and products from the vendors' CSAF advisories.

version not verifiedVersions not verified: read as OSV.dev and the vendors publish them, not checked against each vendor's release list.

Vendor advisories

  • ICSMA-18-144-01 CISA1 known affected

    BeaconMedaes TotalAlert Scroll Medical Air Systems

    TotalAlert Scroll Medical Air Systems: running software Versions 4107600010.23 and prior

Full record on NVD

Fix

  1. Update Beaconmedaes scroll medical air systems firmware to 4107600010.23
  2. Vendor fix CISA ICSMA-18-144-01 · 1 productBeaconMedaes has stated that the vulnerabilities do not compromise either patient health information or compliance with the NFPA 99 standard for healthcare facilities. To address these vulnerabilities, BeaconMedaes has created update 4107600010.24 and recommends that users of the TotalAlert Scroll …

Advisories and patches

Detection

Nuclei template
none so far
ProjectDiscovery
Exploit-DB
no entry
Exploit-DB
Proof-of-concept repositories
none listed
PoC-in-GitHub

Sources

  • NVDCVSS 3.09.8read 2026-10-23
  • NVDCVSS 2.05read 2026-10-23
  • CNAweakness (CWE)CWE-522read 2026-10-23
  • NVDweakness (CWE)CWE-522read 2026-10-23
  • FIRST EPSSEPSS score1.3 %2026-10-22

Technical details

CVSS v3 Vector
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Published
2018-05-24
Last Modified
2026-06-17
NVD status
Modified

Frequently asked questions

What is CVE-2018-7518?

In TotalAlert Web Application in BeaconMedaes Scroll Medical Air Systems prior to v4107600010.23, an attacker with network access to the integrated web server could retrieve default or user defined credentials stored and transmitted in an insecure manner.

Is CVE-2018-7518 actively exploited?

Active exploitation of CVE-2018-7518 has not been confirmed. Its EPSS score was 1.3% on 2026-10-22, the estimated probability of exploitation in the next 30 days.

What is the CVSS score for CVE-2018-7518?

CVE-2018-7518 has a CVSS v3 base score of 9.8 (CRITICAL severity), with vector string CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H.

Hear about the next vulnerability in your software