Description

An Improper Certificate Validation issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump does not validate host certificates, leaving the pump vulnerable to a man-in-the-middle (MITM) attack.

Timeline

dates in UTC
  1. 2018-02-1510:29CVE publishedNVD

Observed activity

isMalicious corpus · updated 2026-10-23 21:37 UTC
Indicators tagged with this CVE, per day0 indicators over 7 days
Block these indicators
  • Firewall list (EDL, text): https://api.ismalicious.com/edl/cve/CVE-2017-12721?format=txt
  • CSV export: https://api.ismalicious.com/edl/cve/CVE-2017-12721?format=csv
  • TAXII collection of exploited CVEs
  • The EDL and CSV take an API key on the Pro plan.

Versions per branch

NVD
Smiths-medical medfusion 4000 wireless syringe infusion pumpversion to verify
BranchAffectedFixed in
1.1= 1.1no fix published
1.5= 1.5no fix published
1.6= 1.6no fix published

as publishedRanges as NVD and OSV publish them; a distribution’s build can be fixed under an older number.

Distribution builds and vendor advisories

Packages from OSV.dev and products from the vendors' CSAF advisories.

version not verifiedVersions not verified: read as OSV.dev and the vendors publish them, not checked against each vendor's release list.

Vendor advisories

  • ICSMA-17-250-02A CISA3 known affected

    Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump Vulnerabilities (Update A)

    Medfusion 4000 Wireless Syringe Infusion Pump: Version 1.1 · Medfusion 4000 Wireless Syringe Infusion Pump: Version 1.6 · Medfusion 4000 Wireless Syringe Infusion Pump: Version 1.5

Full record on NVD

Fix

  1. Vendor fix CISA ICSMA-17-250-02A · 3 productsSmiths Medical has released update Version 1.6.1 to mitigate the vulnerabilities in the Medfusion 4000 Wireless Syringe Infusion Pump.
  2. Vendor fix CISA ICSMA-17-250-02A · 3 productsSmiths Medical recommends users apply the following defensive measures:
  3. Vendor fix CISA ICSMA-17-250-02A · 3 productsFor additional information about the vulnerabilities, proposed measures, or obtaining the product fix, contact Smiths Medical Technical Support at:
  4. Vendor fix CISA ICSMA-17-250-02A · 3 productsTel: +1 (800) 258 5361 or +01 614 210 7300
  5. Vendor fix CISA ICSMA-17-250-02A · 3 productsOnline: https://smiths-medical.custhelp.com/
  6. Vendor fix CISA ICSMA-17-250-02A · 3 productsEmail: info.asd@smiths-medical.com

Detection

Nuclei template
none so far
ProjectDiscovery
Exploit-DB
no entry
Exploit-DB
Proof-of-concept repositories
none listed
PoC-in-GitHub

Sources

  • NVDCVSS 3.05.9read 2026-10-23
  • NVDCVSS 2.04.3read 2026-10-23
  • NVDweakness (CWE)CWE-295read 2026-10-23
  • FIRST EPSSEPSS score0.7 %2026-10-23

Technical details

CVSS v3 Vector
CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N
Published
2018-02-15
Last Modified
2026-06-17
NVD status
Modified

Frequently asked questions

What is CVE-2017-12721?

An Improper Certificate Validation issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump does not validate host certificates, leaving the pump vulnerable to a man-in-the-middle (MITM) attack.

Is CVE-2017-12721 actively exploited?

Active exploitation of CVE-2017-12721 has not been confirmed. Its EPSS score was 0.7% on 2026-10-23, the estimated probability of exploitation in the next 30 days.

What is the CVSS score for CVE-2017-12721?

CVE-2017-12721 has a CVSS v3 base score of 5.9 (MEDIUM severity), with vector string CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N.

Hear about the next vulnerability in your software

Surveillance reads the versions in your stack — a domain, a lockfile or a list — checks each CVE against the ranges its sources publish, and alerts you when one you run is exploited or enters CISA KEV. A version it cannot read stays « version not verified ».

No credit card required · No monthly cap on website lookups · 250 API requests/month