CVE-2016-5085
No exploitation reported by the sources we read.
- Published2016-10-05
- WeaknessCWE-330
Am I affected?
Debian, Ubuntu or Red Hat sometimes backport a fix without changing the upstream version number: such a version reads « to verify », never « affected ».With a domain, we read the version its public headers and pages expose. A distribution build can carry the fix under an older number.
Description
Johnson & Johnson Animas OneTouch Ping devices do not properly generate random numbers, which makes it easier for remote attackers to spoof meters by sniffing the network and then engaging in an authentication handshake.
Timeline
dates in UTC- 2016-10-0510:59CVE publishedNVD
Observed activity
isMalicious corpus · updated 2026-10-23 18:48 UTCBlock these indicators
- Firewall list (EDL, text): https://api.ismalicious.com/edl/cve/CVE-2016-5085?format=txt
- CSV export: https://api.ismalicious.com/edl/cve/CVE-2016-5085?format=csv
- TAXII collection of exploited CVEs
- The EDL and CSV take an API key on the Pro plan.
Versions per branch
NVD| Branch | Affected | Fixed in | Source |
|---|---|---|---|
| — | all versions | no fix published | NVD |
as publishedRanges as NVD and OSV publish them; a distribution’s build can be fixed under an older number.
Distribution builds and vendor advisories
Packages from OSV.dev and products from the vendors' CSAF advisories.
version not verifiedVersions not verified: read as OSV.dev and the vendors publish them, not checked against each vendor's release list.
Vendor advisories
Animas OneTouch Ping Insulin Pump Vulnerabilities
Animas OneTouch Ping insulin pump system: vers:all/*
Fix
- Mitigation from the vendor CISA ICSMA-16-279-01 · 1 productAnimas does not plan to release a firmware update to address the identified vulnerabilities. Animas reports that customer notifications are being sent to patients and HealthCare professionals, which is available on Animas’ web site at the following location: (https://www.animas.com/sites/default/fi…
- Mitigation from the vendor CISA ICSMA-16-279-01 · 1 productAnimas has provided the following compensating controls to help reduce the risk associated with the exploitation of the identified vulnerabilities:
- Mitigation from the vendor CISA ICSMA-16-279-01 · 1 productThe pump’s radio frequency feature can be turned off, which is explained in Chapter 2 of Section III of the OneTouch Ping Owner’s Booklet. However, turning off this feature means that the pump and meter remote will no longer communicate and blood glucose readings will need to be entered manually on…
- Mitigation from the vendor CISA ICSMA-16-279-01 · 1 productIf patients choose to use the meter remote feature, another option for protection is to program the OneTouch Ping pump to limit the amount of bolus insulin that can be delivered. Bolus deliveries can be limited through a number of customizable settings (maximum bolus amount, 2-hour amount, and tota…
- Mitigation from the vendor CISA ICSMA-16-279-01 · 1 productAnimas also suggests turning on the Vibrating Alert feature of the OneTouch Ping system, as described in Chapter 4 of Section I. This notifies the user that a bolus dose is being initiated by the meter remote, which gives the patient the option of canceling the bolus.
- Mitigation from the vendor CISA ICSMA-16-279-01 · 1 productThe bolus delivery alert and the customizable limits on bolus insulin can only be enabled on the pump and cannot be altered by the meter remote. This is also true of basal insulin. Patients can also be reminded that any insulin delivery and the source of the delivery (pump or meter remote) are reco…
Advisories and patches
- community.rapid7.comMitigation
Detection
- Nuclei template
- none so far ProjectDiscovery
- Exploit-DB
- no entry Exploit-DB
- Proof-of-concept repositories
- none listed PoC-in-GitHub
Sources
- NVDCVSS 3.07.5read 2026-10-23
- NVDCVSS 2.07.8read 2026-10-23
- NVDweakness (CWE)CWE-330read 2026-10-23
- FIRST EPSSEPSS score3.9 %2026-10-22
Technical details
- CVSS v3 Vector
- CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N
- Published
- 2016-10-05
- Last Modified
- 2026-06-17
- NVD status
- Modified
Frequently asked questions
What is CVE-2016-5085?
Is CVE-2016-5085 actively exploited?
What is the CVSS score for CVE-2016-5085?
Hear about the next vulnerability in your software
Surveillance reads the versions in your stack — a domain, a lockfile or a list — checks each CVE against the ranges its sources publish, and alerts you when one you run is exploited or enters CISA KEV. A version it cannot read stays « version not verified ».
No credit card required · No monthly cap on website lookups · 250 API requests/month
Go further
- Following what attackers exploit right now? The CISA KEV additions of the month, with their patch deadlines.
- Running this software? Vulnerability monitoring alerts you to new vulnerabilities in your products.
Related vulnerabilities
Same product first, then the same vendor advisory and the same publication week.